INFICON BPG402-Sx ATM to Ultra-High Vacuum Gauge User Manual

Page 33

Advertising
background image

tina46e1-a (2010-03) BPG402.om

33

For transmission of the commands to the gauge, a string (frame) of five bytes is
sent (without <CR>). Byte 1 … 3 form the data string.

Byte no Function

Value

Comment

0

Length of data string

3

set value

1 Data

→ admissible input strings

2 Data

→ admissible input strings

3 Data

→ admissible input strings

4 Check

sum

(from bytes No 1 … 3)

0 … 255


(low byte of sum)

4)

4)

High order bytes are ignored in the check sum.


For commands to the gauge, the following strings are defined (values in decimal
notation):

Command Byte

No

0 1 2 3 4

5)

Set the unit mbar in the display

3

16

142

0

158

Set the unit Torr in the display

3

16

142

1

159

Set the unit Pa in the display

3

16

142

2

160

Power-failure-safe storage of current unit

3

32

2

34

Switch degas on
(switched off automatically after 3 minutes)

3

16

196

1

213

Switch degas off (before 3 minutes)

3

16

196

0

212

Set Emission Control Mode to AUTO

6)

3 16 138 1 155

Set Emission Control Mode to MAN

6)

3 16 138 0 154

Power-failure-safe storage of the
Emission Control Mode

6)

3

32

1

33

Switch emission on

3

64

16

1

81

Switch emission off

3

64

16

0

80

Set Filament Control Mode to AUTO

7)

3 16 211 0 227

Set Filament Control Mode to MAN

7)

3 16 211 1 228

Power-failure-safe storage of the
Filament Control Mode

7)

3

32

13

45

Select filament 1

8)

3 16 210 0 226

Select filament 2

8)

3 16 210 1 227

Power-failure-safe storage of selected filament

8)

3 32 12 – 44

Read

filament

status

3 0 212 – 212

Read

software

version

3 0 209 – 209

Reset

3 64 0 0 64

5)

Only low order byte of sum (high order byte is ignored).

6)

Defines the Emission Control Mode (

→ 29):

AUTO = emission on/off automatically controlled by the gauge
MAN = emission on/off controlled via interfaces.

7)

Defines the Filament Control Mode (

→ 29):

AUTO = Selection of filament automatically controlled by the gauge
MAN = Selection of filament controlled via interfaces.

8)

The "Select filament x" command can be sent any time but is only executed if

the gauge is in the "Emission OFF" state.



4.9.1.2 Input String (Receive)

Format of the input string

Admissible input strings