For cable lengths up to 5 m (0.34 mm

2

conductor cross-section) the out-

put signal can be measured directly between the positive signal output
(Pin 2) and supply common (Pin 5). At greater cable lengths, differential
measurement between signal output (Pin 2) and signal common (Pin 12)
is recommended.

Reassemble the cable connector.

On the other cable end, terminate the cable according to the requirements

of the gauge controller you are using.

Plug the sensor connector into
the gauge and secure it with the
locking screws.

Connect the other end of the sensor cable to the connector of the instru-

ment or gauge controller you are using.


For operating BPG402-SD via DeviceNet, an interface cable conforming to the
DeviceNet standard is required.
If no such cable is available, make one according to the following indications.

A shielded special 5 conductor cable conforming to the DeviceNet standard has to
be used (

→ [4], [6]).

Make the DeviceNet cable according to the following indications.

3

1

4

2

5


Micro-Style, 5 pins,
(DeviceNet)
female, soldering side

Pin

1 Drain

Pin 2 Supply (DeviceNet interface only)

+24 VDC

Pin 3 Supply common (DeviceNet interface only)

GND

Pin

4 CAN_H

Pin

5 CAN_L

3.2.2.2 Making a DeviceNet

Interface Cable
(BPG402-SD)

Cable type

Procedure