tina46e1-a (2010-03) BPG402.om

9

BPG402-S, -SL

BPG402-SD, -SP

Adjustment range

1 ("SETPOINT")

2 ("SETPOINT A, B")

1×10

-9

mbar … 100 mbar

Setpoints adjustable via potentiometers,
one floating, normally open relay contact
per setpoint (

→ 19, 20, 37).

(Adjusting the setpoints via field bus
→ corresponding bus section)

Hysteresis

10% of the

threshold value

Relaiskontaktbelastung

≤30 VDC, ≤0.5 ADC

Data rate

Data format



Connections (sensor cable connector)

TxD (Transmit Data)
RxD (Receive Data)
GND

9600 Baud

binary
8 data bits
one stop bit
no parity bit
no handshake


Pin 13
Pin 14
Pin 5

Function and communication protocol of the RS232C interface

→ 30

Fieldbus name

DeviceNet

Standard applied

→ [6]

Communication protocol, data format

→ [1], [4]

Interface, physical

CAN bus

Data rate
(adjustable via "RATE" switch)

125 kBaud
250 kBaud
500 kBaud (default)
"P" (125 kBaud, 250 kBaud, 500 kBaud
programmable via DeviceNet)
(

→ [1])

Node address (MAC ID)
(Adjustable via "ADDRESS", "MSD",
"LSD" switches)

0 … 63

dec

(default = 63

dec

)

"P" (0 … 63 programmable via
DeviceNet,

→ [1])

DeviceNet connector

Micro-Style, 5 pins, male

Cable

shielded, special DeviceNet cable,
5 conductors (

→ 21 and [4])

Cable length, system wiring

according to DeviceNet specifications
(

→ [6], [4])

Switching functions

RS232C interface

DeviceNet interface
(BPG402-SD)