INFICON BPG402-Sx ATM to Ultra-High Vacuum Gauge User Manual

Page 11

Advertising
background image

tina46e1-a (2010-03) BPG402.om

11

For reasons of compatibility, the expression "sensor cable" is used for all
BPG402 versions in this document, although the pressure reading of the
gauges with fieldbus interface (BPG402-SD and BPG402-SP) is
normally transmitted via the corresponding bus.

Electrical connection

BPG402-S, -SL
BPG402-SD, -SP

D-Sub,15 pins, male
→ 19

→ 20

Sensor cable

shielded, number of conductors de-
pending on the functions used,
max. 15 conductors plus shielding

Cable length (supply voltage 24 VDC

1)

)

Analog and fieldbus operation

For operation with RS232C interface


≤35 m, 0.25 mm²/conductor

≤50 m, 0.34 mm²/conductor

≤100 m, 1.0 mm²/conductor
≤30 m

Materials exposed to vacuum

Housing, supports, screens
Feedthroughs
Insulator
Cathode
Cathode holder
Pirani element


stainless steel
NiFe, nickel plated
glass
iridium, yttrium oxide (Y

2

O

3

)

molybdenum, platinum
tungsten, copper

Internal volume

DN 25 ISO-KF
DN 40 CF-R

24 cm

3

34 cm

3

Pressure max.

2 bar (absolute)


Admissible temperatures

Storage

Operation

Bakeout
Long

tube

-20 … 170 °C

0 … 150 °C

+ 80 °C

2)

+150 °C

2)

Relative humidity

Year's mean
During 60 days


≤65 (no condensation)
85% (no condensation)

Use indoors

only

altitude up to 2000 m NN

Mounting orientation

any

Type of protection

IP 30

1)

Measured at gauge

connector (consider the voltage drop as function of the sensor cable

length).

2)

Flange temperature, electronics unit removed, horizontally mounted.

Electrical connection

Materials used

Ambiance

Advertising