2 using the optional baffle – INFICON BPG402-Sx ATM to Ultra-High Vacuum Gauge User Manual

Page 15

Advertising
background image

tina46e1-a (2010-03) BPG402.om

15

Place the electronics unit (2) on the sensor (3) (be careful to correctly align

the pins and notch (4)).

4

3

2

Slide the electronics unit in to the mechanical stop and lock it with the hexa-

gon socket set screw.


In severely contaminating processes and to protect measurement electrodes opti-
cally against light and fast charged particles, replacement of the built-in grid by the
optional baffle (

→ 45) is recommended.

Gauge deinstalled

("Deinstallation"

→ 39).


• Baffle

(

→ 45)

• Pointed

tweezers

• Pin (e.g. pencil)
• Screwdriver No 1

Carefully remove the grid with tweezers.

Installing the electronics unit

3.1.2 Using the Optional Baffle

Precondition

Required tools / material

Installation