1 removing and installing the electronics unit – INFICON BPG402-Sx ATM to Ultra-High Vacuum Gauge User Manual
Page 14
Advertising
14
tina46e1-a (2010-03) BPG402.om
Remove the protective lid and install the gauge to the vacuum system.
Seal with centering ring
Protective lid
(keep it)
Clamp
• Allen wrench, AF 2.5
Unscrew the hexagon socket set screw (1) on the side of the electronics
unit (2).
2
1
Remove the electronics unit without twisting it.
Procedure
3.1.1 Removing and Installing
the Electronics Unit
Required tools / material
Removing the electronics unit