12 switching functions – INFICON BPG402-Sx ATM to Ultra-High Vacuum Gauge User Manual

Page 37

Advertising
background image

tina46e1-a (2010-03) BPG402.om

37

The BPG402-S, SL have one, the gauges BPG402-SD and BPG402-SP have two
independent, manually adjustable switching functions. Each switching function has
a floating, normally open relay contact. The relay contacts are accessible at the
sensor cable connector (

→ 19, 20).

The threshold values of the switching functions can be set within the pressure
range 1×10

-9

mbar … 100 mbar via potentiometers "SETPOINT" (BPG402-S, SL)

or "SETPOINT A" and "SETPOINT B" (BPG402-SD, SP).

The following rule applies:

U

Threshold

= 0.75 × (log p

Setpoint

– c) + 7.75

Where constant c is pressure unit dependent (

→ Appendix A).

Me

asu

red va

lue

Off

On

Switching function

Measuring Signal
(Pressure p)

(Setpoint A, B)

Time t

Off

U

Threshold

Hysteresis
10% U

Threshold

The hysteresis of the switching functions is 10% of the threshold setting.


4.12 Switching

Functions