4.

Initial Operation

4.3 Electrical connection
4.3.1 Checking the voltage selector setting

Fig. 13

The voltage selector is located in a shell (3) above

the power switch (2) at the rear of the instrument.

The actual selected voltage can be seen in the

window (1).

• Check if the setting showing in the window

(1) corresponds to the nominal voltage in your

laboratory.

• Remove the adhesive tape.

If the setting is correct proceed at

4.3.3.

If the setting does not correspond to the nominal

voltage in your laboratory, the voltage selector

MUST be adjusted to match your laboratory's

electrical setting prior to connecting the instru-

ment to the mains.

230

3

230 V

2

1

Important! The voltage selector is factory preset.

Before connecting the instrument to the power supply, make sure this setting is correct for your

laboratory's electrical connection values.

The socket for connecting the power cable is sealed with an adhesive tape indicating the

factory preset voltage for your particular instrument.

Connecting the instrument to the power with the voltage selector set at a wrong value can

cause severe damage to the instrument!