Leica TP1020

4. Instrument Setup

Fig. 14

5

4

2

4.3.2 Adjusting the voltage selector

• Insert the small screwdriver (4) into the smaller

one of the two notches at the lower end of the

lock (5) using the screwdriver as a lever.

• Remove the shell (3).
• Remove the voltage selector (6) from the shell

(3) and reinsert it so that the correct voltage

selection can be seen in the window (1) of the

shell.

• Insert the voltage selector shell (3) together

with the voltage selector (6) and fuses (7) back

into the receptacle in the instrument and press

lightly until it locks.

• Check again if the right setting shows in the

window (1).

Fig. 15

1

6

7

3

When adjusting the voltage selector

setting, the instrument must be discon-

nected from the mains.