10. Warning and Error Codes – Troubleshooting

List of error codes (continued)

Error code "E:11" Loss of process data.

- Acknowledge the error message by

pressing any key; "ABORT" message,

the process is canceled

- Restart processing cycle.

- If the same problem recurs, call

Technical Service.

Error code "E:12"

Loss of program settings.

- Acknowledge the error message by

pressing any key; "ABORT" message,

the process is canceled

- Restart processing cycle.

- If the same problem recurs, call

Technical Service.

Error code "E:13"

Erroneous data entry for delayed start

function (e.g. desired starting time

would be earlier than current real time.)

- Correct erroneous data entry.