Appendix, 1 modifying the instrument configuration – Leica Biosystems TP1020 User Manual

Page 90

Advertising
background image

90

Instructions for Use V 2.9 RevC, English – 12/2013

15.1 Modifying the instrument configuration

+

The instrument has a factory-set standard con-

figuration which can be modified by the user if

desired.

Modifications can be done in the configuration

menu, which consists of 4 different menu items:

- Signal level 1 - 4

- 3. Wax bath?

- ProgPreset?

- SystemReset?

Editing the configuration menu
• Switch the instrument off at the main power

switch.

• Turn the power switch back on. While the

display still reads "TP1020 V x.xx" press PROG

for a short while.

The first configuration menu item is displayed.

Modifying the configuration
Signal level = Setting of the desired type and

frequency of sound signal

Level 1 = sound signal in case of power failure

and malfunctions.

Level 2 = level 1 + sound signal in case of inter-

rupt ion, warning code and end of

program.

Level 3 = level 2 + sound signal when confirm-

ing warning codes and error codes.

Level 4 = level 3 and sound signal every time a

button is pressed.

15. Appendix

For this menu item four different settings

can be selected. Setting 4 is factory

preset.

Setting 1 is the lowest level. The next

higher level always contains all the

parameters of the next preceding level

plus the additional signals as listed on

this page.

Advertising