4 pausing a process, Pausing a process – Leica Biosystems TP1020 User Manual

Page 64

Advertising
background image

64

Instructions for Use V 2.9 RevC, English – 12/2013

9.

Operating the Instrument in the Automatic Processing Mode

In the example shown here end of processing

would be the next day at 10.38 AM.

• To quit the display indication, press CLOCK

again.

9.4 Pausing a process

Automatic specimen processing can be inter-

rupted and resumed later, for example to reload

specimens.

• To interrupt the program, press the PAUSE/

CONT button.

"PAUSE" is displayed.

When in "PAUSE", the buttons for manual opera-

tion become functional, e.g. to lift the tissue bas-

ket out of a station to add specimens. The car-

ousel can also be rotated to move the basket to

any processing station - see

chapter 8 "Operating

the Instrument in the Manual Processing Mode".

Editing and modifying programs while a processing cycle is in progress (continued)

While a run is paused, "PAUSE" will be displayed continuously. In addition, 5 minutes after last

pressing a button, a double sound signal will remind the user that processing is still paused.

The sound signal will recur in 5 minute intervals until processing is resumed.