9. Operating the Instrument in the Automatic Processing Mode

Delayed start (continued)

Edit and change starting time (when using the delayed starting function)

+

• Press CLOCK again.

• Check the selected starting time and if neces-

sary change it as appropriate to achieve the

desired end or run time.

• To activate the delay function press START.

• To override the warning codes and start pro-

cessing, press KEY and START simultaneously.

The program will be started at the selected delay

time.

• To quit the display indication, press CLOCK

again.

The warning codes W:04 - W:06 (see Chapter 7) are displayed for programs of less than 8 hours

duration from start until the basket reaches the first wax bath. Check if the available amount of

time is sufficient for the paraffin to liquefy completely. If this is not ensured, fill the wax baths

with liquid paraffin.