10. Warning and Error Codes – Troubleshooting

10.1 Warning codes

The warning codes W:01 - W:06 may be dis-

played.

W:01, W:02, W 03 - paraffin in station No. 10, 11,

12 is still solid.

W:04, W:05, W:06 - paraffin in station No. 10, 11,

12 may still be solid when

basket No. 1 arrives.

When these warning codes are displayed, a short

sound signal (very short sound - very short sound)

is emitted and repeated in 5 minute intervals.

W:01 - W:03 are displayed by the processing cy-

cle upon occurrence, e.g. when the basket is due

to be immersed into a wax bath as programmed.
W:04 - W:06 are displayed immediately when

starting an automatic processing cycle of less

than 8 hours duration from start until the basket

reaches the first wax bath.

• To quit warning codes W:01 - W:03 press any

key.

• To lower or lift the tissue basket press KEY and

ARROW UP or ARROW DOWN simultaneously.

+

+

Or

Solid paraffin takes several hours to

melt.

Please keep this in mind and make

sure to switch on the instrument early

enough so the paraffin will be molten

and ready for use when necessary.