Instructions for Use V 2.9 RevC, English – 12/2013

6.

Programming the Instrument

6.2.4 Selecting the starting station

• Press ARROW RIGHT to move the cursor to

the next entry position.

The cursor blinks at the station number data entry

position.

• Select the desired station number by pressing

PLUS or MINUS.

6.2.5 Activating the vacuum function

(instruments with vacuum function only)

• Press ARROW RIGHT to move the cursor to

the next entry position.

The cursor blinks at the two positions next to the

"V" for vacuum function.

• To activate the vacuum, press PLUS.

• To deactivate the vacuum, press MINUS.

The Leica vacuum accessories for en-

hanced tissue infiltrations may ONLY

be used with aluminum containers.

Glass beakers used in combination

with the vacuum accessories have a

high hazard potential when damaged!