Modifying the instrument configuration (continued)

• For the next menu item, press the CURSOR

button.

• To change the setting, press the PLUS / MINUS

keys.

The next standard setting is displayed.

3. Wax bath = do you wish to use a third wax

bath?

• To change the setting, press the PLUS / MINUS

keys.

• For the next menu item, press the CURSOR

button.

The next standard setting is displayed.

ProgPreset? = Set programs no. 6 - 9

to factory preselected standard

values?

In station no. 10 you can install a third wax bath instead of the reagent station.

If you do so, you have to modify the configuration menu accordingly; otherwise the instrument

will not function properly. If you change from a three wax baths operation mode back to two

wax baths, you have to go back and readjust the configuration menu.