7.

Preparations for Tissue Processing

7.2 Filling the wax baths

Fig. 21

1

• Use a screw driver to turn the setting screw

(1) to the desired value.

If you find that the paraffin does not melt com-

pletely after lowering the working temperature,

slightly readjust again.

Altering the standard working temperature

The heated wax baths may only be used with paraffin. Under no

circumstances may they be filled with solvents. When solvents

heat, a highly explosive mixture builds up!

Caution! The interior containers of the wax baths become very hot

when the heating function is activated! Do not touch the gray upper

rim of the containers with your hands! Burn hazard!

Caution when handling hot paraffin! Burn hazard!

Factory-set standard working temperature is 65 °C (70 °C on the spe-

cial wax bath model that is resistant to chloroform). When working

with paraffin that has a melting point below 58 °C, the instrument

working temperature can be readjusted with the corresponding

setting screw.