Leica TP1020

12. Maintenance

12.2 Wax bath
12.2.1 Altering the standard working temperature

Fig. 24

• Use a screw driver to turn the setting screw

(1) to the desired value.

If you find that the paraffin does not melt com-

pletely after lowering the working temperature,

slightly readjust again.

1

Factory-set standard working temperature is 65 °C. Wax bath heating will automatically switch

on when the actual temperature of the paraffin is 5 °C below set working temperature.

When working with paraffin that has a melting point below 58 °C, the instrument working

temperature can be readjusted with the corresponding setting screw.