3 inserting the tissue basket, Inserting the tissue basket, Preparations for tissue processing – Leica Biosystems TP1020 User Manual

Page 52

Advertising
background image

52

Instructions for Use V 2.9 RevC, English – 12/2013

7.

Preparations for Tissue Processing

7.3 Inserting the tissue basket

Fig. 22

• Fill the tissue cassettes or capsules into

the tissue basket.

• To hook in the tissue basket(s), the

basket holder(s) should not be located

over a wax bath.

• Lift the carousel in the manual process-

ing mode.

• Rotate the carousel so the basket

holder (1) is located above the starting

container for basket 1.

• Hook the tissue basket (2) into the

basket holder as shown.

• Lower the basket into the starting con-

tainer in the manual processing mode

or start an automatic processing cycle.

1

2

When working in the manual mode, all carousel movements are activated

through the corresponding function keys on the control panel.

Use caution when lowering the carousel! Keep your fingers out of

the space between the container lid and the upper rim of the con-

tainer!

The carousel may not be rotated manually! Severe damage will

result from doing so!