Safety – Leica Biosystems ASP300 S User Manual

Page 10

Advertising
background image

11

Instructions for Use, V 1.5 RevD – 03/2013

2. Safety

Transport and Installation

Warnings – Markings on the instrument itself

Markings on the instrument showing the warning triangle indicate that the correct operat-

ing instructions (as defined in these Instructions for Use) must be followed when operating or

replacing the item marked. Failure to adhere to these instructions may result in an accident,

personal injury, damage to the instrument or accessory equipment.

Some instrument surfaces become hot during operation. They are marked with this

warning label. Touching these surfaces may cause burns.

Once unpacked, the instrument may be transported only in an upright position.

Follow the unpacking instructions carefully to avoid damage to the instrument!

Prior to every transport in which it is possible for the instrument to be shaken, tilted, or lifted,

it must be cleaned for transport – otherwise the interior of the instrument can be severely dam-

aged.

Plug the instrument only into a grounded power socket. Do not interfere with the grounding

function by using an extension cord without a ground wire.

Make sure to observe the voltage settings!

The set voltage CANNOT be changed by the user.

Severe damage may occur if the instrument is connected to a power supply voltage other than

that to which it was originally set.

The installation location must be well-ventilated; there should be no ignition sources there of

any kind. The chemicals to be used in the Leica ASP300 S are both flammable and noxious.

Do not operate the instrument in rooms with explosion hazard.

Extreme temperature fluctuations between storage facility and setup site as well as high

humidity may cause condensation to form. If this is the case, wait at least two hours before

switching on the instrument. Failure to comply with this may cause damage to the instrument.

2.2 Warnings