1 cycle for changing reagents, Instrument components and specifications – Leica Biosystems ASP300 S User Manual

Page 22

Advertising
background image

23

Instructions for Use, V 1.5 RevD – 03/2013

3. Instrument Components and Specifications

3.6.1 Cycle for changing reagents

Reagent

Week 1

Week 2

Week 3

Week 4

Formalin

At least 3 times a

week

At least 3 times a

week

At least 3 times a

week

At least 3 times a

week

Formalin

At least 3 times a

week

At least 3 times a

week

At least 3 times a

week

At least 3 times a

week

70 %

Daily

Daily

Daily

Daily

80 %

Once a week

Once a week

Once a week

Once a week

95 %

Once a week

---

Once a week

---

95 %

---

Once a week

---

Once a week

100 %

Once a week

---

Once a week

---

100 %

---

Once a week

---

Once a week

Xylene

Once a week

---

Once a week

---

Xylene

---

Once a week

---

Once a week

Paraffin 1

Run paraffin clean-

ing program* daily.

Replace the paraf-

fin after 6 cleaning

cycles.

Run the paraffin

cleaning program*

once a week.

Run the paraffin

cleaning program*

twice a week.

Run paraffin clean-

ing program* daily.

Replace the paraf-

fin after 6 cleaning

cycles.

Paraffin 2

Run the paraffin

cleaning program*

twice a week.

Run paraffin clean-

ing program* daily.

Replace the paraf-

fin after 6 cleaning

cycles.

Run the paraffin

cleaning program*

once a week.

Run the paraffin

cleaning program*

twice a week.

Paraffin 3

Run the paraffin

cleaning program*

once a week.

Run the paraffin

cleaning program*

twice a week.

Run paraffin clean-

ing program* daily.

Replace the paraf-

fin after 6 cleaning

cycles.

Run the paraffin

cleaning program*

once a week.

Cleaning xylene

Once a week

Once a week

Once a week

Once a week

Cleaning alcohol

Once a week

Once a week

Once a week

Once a week

Cleaning Water

Once a week

Once a week

Once a week

Once a week

* Paraffin cleaning program,

see Chap. 7.1.2