7.2.2 Periodic cleaning and maintenance

Cleaning the reagent containers
• Empty and clean the reagent containers.
Using a bottle brush and a laboratory detergent in warm water.

7. Cleaning and maintenance

Never clean reagent containers in an automatic dishwasher.

The reagent containers are NOT dishwasher-proof!

The reagent containers must be properly engaged in the home po-

sition in the connection manifolds at the rear inner wall of the re-

agent module. Failure to correctly plug reagent containers into the

manifold will cause an interruption to the processing run and may

result in spilling of reagents.

• While the reagent bottles are outside of the reagent cabinet, wipe the

stainless steel inner walls of the reagent cabinet with a damp cloth

moistened with a mild detergent.

Additional cleaning and maintenance tasks to be carried out periodically

are listed in the table on the following pages.

• Refill and reconnect the bottles once cleaned.
Make sure the bottle lids are tight and the bottles are properly seated in

their home position at the rear of the reagent module.