5. Operation

Deleting reagents
• Highlight the reagent to be deleted in the SET UP REAGENTS AND

WARNING THRESHOLDS screen.

• Press CLEAR.
• Press YES in the screen to confirm the reagent is deleted.

Fig. 42

Please remember that a reagent which is already used in a program

cannot be deleted.

Upon activation of the RMS, all warning thresholds for the selected

reagent are automatically reset to "0".

Adding new reagents to stations
Go to SMART SCREEN to remote-fill the reagent container from an external

bulk container
or
fill the reagent container manually.
After filling a bottle manually the bottle must be defined as full.
To do so, follow these steps:
• Press the MENU icon to call up the MENU FUNCTIONS window.
• Press the REAGENT STATUS button.
• In the table, highlight the reagent that has been manually filled.
• Press the SET AS FULL button symbol to set the status of the station to

"full".