4 reagent status, Reagent status, Operation – Leica Biosystems ASP300 S User Manual

Page 66

Advertising
background image

67

Instructions for Use, V 1.5 RevD – 03/2013

5.4 Reagent status

This window (

Fig. 64

) is used to display and update the current reagent

status of the stations. In addition, reagent stations can be marked as

full or empty here.

Fig. 75

The following is displayed:
• The current age of each reagent
• The current status of each station
• Criteria that trigger warning mes-

sages (Fields colored in red)

To change the status of a reagent,

highlight the corresponding row

and press the button symbol for the

respective action.

In addition, you can switch directly

to the REAGENTS and STATIONS

windows to make current changes.

Reagent warning messages
If the RMS is active, warning messages are out-

put if reagents are used for too long.
The output of the warning messages can take

place after up to 3 criteria for normal reagents

and after up to 6 criteria for paraffins.
Warning messages are displayed at the end of a

cleaning program. Then, you can switch directly

to the REAGENT STATUS display.
Warning messages also appear (with a yellow

background) in the windows in which programs

are started.

The criteria for warning messages
For normal reagents, it is the number of the

following since the last reagent change:
• Edited blocks

• Run programs

• Days passed

For paraffins, warning messages can also be

output according to the number of the following

since the last paraffin cleaning:
• Edited blocks

• Run programs

• Days passed

5. Operation