2 replacing the paraffin, Operation – Leica Biosystems ASP300 S User Manual

Page 57

Advertising
background image

58

Leica ASP300 S

5.2.2 Replacing the paraffin

Draining the paraffin
• Connect the paraffin drain hose (36) to the

spout (17) of the paraffin drain at the front of

the instrument (

Fig. 55

) and into a catch tank.

Fig. 55

• Select the paraffin reservoir to be emptied and

the receptacle (External,

Fig. 56

).

Then, press STATION DRAIN....

• This is followed by multiple security prompts

that you have to confirm individually in order

to continue.

17

36

From FAVORITES, call up the SMART SCREEN

screen. (see previous chapter)

+

Fig. 56

5. Operation

When connecting the hose, be sure to

press it onto the O-rings of the drain

opening as far as it will go.

The paraffin drain hose must be

securely hooked into the external

receptacle and remain there through

the entire drainage process.

When drainage is finished, the hose is

cleaned automatically using air.

Do not remove the hose from the exter-

nal receptacle until this cleaning step

is complete.