2 connecting an uninterruptible power supply (ups), Initial operation – Leica Biosystems ASP300 S User Manual

Page 25

Advertising
background image

26

Leica ASP300 S

• Connect the ASP to a grounded socket using

the power cable (37).

• Using the jumper cable (38), connect the output

of the main voltage supply to the UPS input.

• Using the jumper cable (38) connect the input

of the electronics section (40) to one of the

UPS outputs.

• Switch the instrument on at the main switch.
• Start the UPS.

4.2.2 Connecting an uninterruptible power supply (UPS)

38

40

37

38a

Fig. 4a is provided as an example only

and shows the correct connection

of the ASP300 S to an uninterruptible

power supply (UPS).

Fig. 4a

An uninterruptible power supply (UPS)

protects machines and instruments

against malfunctions in the power

supply.

Leica recommends using an active

tracking UPS (with an output power of

1000-1500 VA) to protect the instrument

and the specimens from temporary

power failure, voltage spikes, under-

voltages and excess voltages.

4. Initial Operation