7.1.2 Paraffin cleaning

It is possible to clean used paraffin of solvent

residues (xylene) in the Leica ASP300 S.

To do so, select the paraffin bath for which the

paraffin should be cleaned in the CLEANING

PROGRAMS menu.
If the last reagent in the retort is not compatible

with paraffin, carry out a retort cleaning first.

The following steps are taken during the paraffin

cleaning:
• The paraffin is pumped into the retort.
• There it is heated to the highest possible tem-

perature.

• The retort is put under vacuum, causing the

solvent vapors to be drawn out of the paraffin.

The duration of the paraffin cleaning process is

programmable.
To do so, follow these steps:
• In the PROGRAMS menu, select the PARAFFIN

CLEANING line and press the EDIT button.

• In the entry window, specify the duration for

the cleaning program.

The recommended time frame is one hour.
• Confirm the time specification with OK.