PROGRAM OPTIONS
Select the options with which programs are executed here. Press a button

to change the associated value.
STATION SEQUENCE:

By Age or Sequential

By Age -

RMS is activated, reagents will automatically be

used in order of increasing cleanliness.

Sequential -

reagents will be used in sequential order of stations.

WAX BATH ORDER:

Auto or 1. ; 2. ; 3.

Can only be activated if STATION SEQUENCE is set to "Sequential" –

otherwise the order of baths will be managed by the RMS.

PROMPT FOR NUM BLOCKS: Enabled or Disabled
Enabled -

When starting a program, the actual number of

blocks must be entered. This is recommended with

the RMS being enabled.

Disabled -

Number can be entered, but does not have to be.

DATE / TIME
Be sure to verify that the date and time entered do actually correspond

to the local time/date, as this will ensure that all programs are carried out

correctly.
Press the buttons to access the entry windows where settings or correc-

tions can be made, if necessary.

Date and time setup screen