3 checklist for preventive maintenance, Checklist for preventive maintenance, Cleaning and maintenance – Leica Biosystems ASP300 S User Manual
Page 82
83
Instructions for Use, V 1.5 RevD – 03/2013
7.3 Checklist for preventive maintenance
7. Cleaning and maintenance
Step
Daily
Weekly
Monthly
every
3 months
Lubricate the O-rings of the reagent
container and check them for damage.
√
√
*
* During
remote filling
and draining
After exiting the retort cleaning program,
wipe the retort and cover dry.
√
Clean the outer surfaces of the instrument
with a soft cloth and a very small amount
of xylene.
√
Check the retort filter screen for tissue or
paraffin residue.
√
Remove paraffin residue from the interior
surface of the paraffin reservoir cover.
√
Check the retort lid seal and clean it
if necessary.
√
Check the paraffin reservoir lid seal and
clean it if necessary.
√
Check the paraffin level and refill if
necessary.
√
Check the fill levels of the reagent
containers.
√