Cleaning and maintenance – Leica Biosystems ASP300 S User Manual

Page 73

Advertising
background image

74

Leica ASP300 S

Editing steps

• To edit a program step highlight the corresponding line and press the

respective table header.

• To disable a step, set the number of cycles to "0". Only cleaning water

steps can be disabled.

The following characteristics of each step can be edited:

Reagent name – only those reagents that are compatible with the re-

agent station being edited will be displayed.

Number of cycles – the duration displayed is calculated based on the

number of cycles.

Retort temperature – temperatures from 50 to 65 °C can be set.

7. Cleaning and Maintenance

The finish times of clean programs cannot be edited. Program dura-

tion is calculated automatically based on the number of cycles set.

The station 13-ext can be assigned exclusively to reagents from the

cleaning water group. It may only be used for cleaning programs.

Station 13-ext
• The designation "13-ext" is used for an additional station that can be used

in place of the standard station 13.

• Station no. 13 always contains cleaning water. For some applications it

is desirable to use a cleaning water/agent mixture between the cleaning

alcohol and the cleaning water.

• If a program contains the step "Station 13-ext", the operator is prompt-

ed to insert the reagent container "13-ext" at the required point in time

during the cleaning program, then to remove it once it is no longer used.