7 switching the instrument on, Switching the instrument on, Initial operation – Leica Biosystems ASP300 S User Manual

Page 31

Advertising
background image

32

Leica ASP300 S

4.7 Switching the instrument on

• After being switched on, the instrument will

take a few minutes to initialize. See opposite for

corresponding touchscreen display (

Fig. 13

).

• If the retort is not locked, a signal tone sounds for

approx. 10 sec. during initialization of the instru-

ment. It is not possible to lock the retort during this

time.

• If the retort is locked, no signal tone sounds. A sig-

nal tone sounds for approx. 10 seconds (waiting

time) when attempting to open the retort. The retort

cannot be opened during this time.

• After that, the FAVORITES screen will be displayed

(

Fig. 14

).

Screen saver
• A screen saver will turn off the screen display if no

key has been pressed for a (user-programmable)

time. Press any part of the touchscreen to restore

the screen.

After having been restored, the on-screen func-

tions will be nonoperational for a few seconds to

avoid accidental activation of any keys.

Fig. 14

Fig. 13

4. Initial Operation

The instrument MUST be connected to a grounded power socket.

For additional electrical fuse protection, we recommend connecting the Leica ASP300 S to a

socket with a residual current circuit breaker.

• Connect the power cable to the power socket. If applicable, switch on the switch for

the power socket.

• To switch on the instrument, press in the ON/OFF switch on the rear panel of the

instrument (

Item 42 in Fig. 4

) switching on (ON).

• Switch on the ON/STOP switch on the right of the instrument (ON).

The ON/OFF switch and the side ON/STOP switch should remain switched on at all times to

keep the heater running for the wax baths.

The ON/STOP switch can be used in emergencies to stop the instrument during while a proce-

dure is in progress.

Advertising