4.7 Switching the instrument on

• After being switched on, the instrument will

take a few minutes to initialize. See opposite for

corresponding touchscreen display (

Fig. 13

).

• If the retort is not locked, a signal tone sounds for

approx. 10 sec. during initialization of the instru-

ment. It is not possible to lock the retort during this

time.

• If the retort is locked, no signal tone sounds. A sig-

nal tone sounds for approx. 10 seconds (waiting

time) when attempting to open the retort. The retort

cannot be opened during this time.

• After that, the FAVORITES screen will be displayed

(

Fig. 14

).

Screen saver
• A screen saver will turn off the screen display if no

key has been pressed for a (user-programmable)

time. Press any part of the touchscreen to restore

the screen.

After having been restored, the on-screen func-

tions will be nonoperational for a few seconds to

avoid accidental activation of any keys.

Fig. 14

Fig. 13

4. Initial Operation

The instrument MUST be connected to a grounded power socket.

For additional electrical fuse protection, we recommend connecting the Leica ASP300 S to a

socket with a residual current circuit breaker.

Connect the power cable to the power socket. If applicable, switch on the switch for

the power socket.

To switch on the instrument, press in the ON/OFF switch on the rear panel of the

instrument (

Item 42 in Fig. 4

) switching on (ON).

Switch on the ON/STOP switch on the right of the instrument (ON).

The ON/OFF switch and the side ON/STOP switch should remain switched on at all times to

keep the heater running for the wax baths.

The ON/STOP switch can be used in emergencies to stop the instrument during while a proce-

dure is in progress.