Replacing the active carbon filter

• The life of the active carbon filter will depend

on the reagent types used and the frequency

of vacuum cycles.

• The filter (13) should be replaced at least every

3 months (

Fig. 84

).

After replacing the filter, open the SMART

MONITOR menu to reset the filter age to "0".

For additional information, refer to

Chap. 5.5

• We recommend a smart clean prior to chang-

ing the filter.

Fig. 84

Draining the condensate container

• Remove, empty, and reinsert the blue conden-

sate container (

Fig. 83

).

Fig. 83

Filter label

7. Cleaning and Maintenance

14

13

Dispose of waste solvents with care

according to local regulations and the

waste management policy of the com-

pany or institution.

When inserting the new filter, make

sure that it is inserted with the correct

side up, as specified on the label on

the front side of the filter. The arrow

must point upwards.