The RUN OPTIONS function displays the options

with which the current program is started.

LEVEL TEST enabled:

The lower level sensor in the retort is acti-

vated. The program is stopped with an error

message if the level is not reached.

WAVE MOTION activated:

During each infiltration step, the reagent is

periodically pumped from the bottle to the re-

tort and back to achieve a more evenly mixed

blend. The first pump step begins after 16 min,

afterwards regularly after 12 min.

PRE-TEST activated:

After the program is started, the retort is filled

and drained with the first reagent in the infil-

tration program (to test whether all lines and

valves are unobstructed).

STIRRER activated:

The magnetic stirrer is on.

PRINTER, if activated:

At the end of the program, a complete run

log covering the entire program execution is

output at the printer.

Fig. 66

Important!

The options set apply for ALL programs!

All run options can be deactivated or activated according to requirements while the program is

running.

The printer option will be displayed only if a printer has been configured for the instrument.