8

Instructions for Use, V 1.8 RevC – 07/2013

2. Safety

Hazards – servicing and cleaning

Safety instructions – working with the instrument

• Take care when handling microtome knives and disposable blades. The cutting edge is

extremely sharp and can cause serious injuries!

• Always remove the knife / blade before detaching the knife holder from the instrument.

Always put the knives back into the knife case when not in use!

• Never place a knife anywhere with the cutting edge facing upwards and never try to catch a

falling knife!

• Always clamp the specimen block BEFORE clamping the knife or blade.

• Lock the handwheel and cover the knife edge with the knife guard prior to any manipula-

tion of knife or specimen clamp, prior to changing the specimen block and during all work

breaks!

• ALWAYS turn the handwheel clockwise; otherwise, the brake will not work properly.

• Always take appropriate safety precautions when sectioning brittle specimens! Specimens

may splinter!

• Ensure that liquids do not enter the interior of the instrument during work!

• Do not attempt to clamp, approach or orient the specimen during the retraction phase.

• If a block is oriented during retraction, the block will advance by the retraction value PLUS

the selected section thickness before the next section. This may cause damage to both speci-

men and knife!

• The handle of the handwheel must always be centered while in motorized sectioning mode.

Do not touch the handwheel while it is running – there is a danger of injury from the hand-

wheel lock.

• Prior to sectioning, check that the specimen is securely clamped in the specimen clamp –

failure to observe this poses the risk of damaging the specimen.

• Only authorized and qualified service personnel may access the internal components of the

instrument for service and repair!

• Before each cleaning, switch off the instrument, disconnect the power plug, remove the knife

holder completely and clean it separately.

• Always remove the blade before detaching the knife holder from the instrument.

• Lock the handwheel before each cleaning!

• Do not use any solvents containing acetone or xylene for cleaning!

• Ensure that liquids do not enter the interior of the instrument during cleaning!

• Do not turn the instrument on before it is completely dry!

• When using detergents please comply with the safety precautions of the manufacturer.

• Turn the instrument off with the power switch and disconnect the power plug before re-

placing the fuses! Only use fuses of the same specification! For fuse specifications, refer to

Chapter 3.3 – "Technical Data"

.