3 gerät ölen, 3 lubricating the instrument, Cleaning and maintenance – Leica Biosystems RM2265 User Manual

Page 82

Advertising
background image

80

Instructions for Use, V 1.8 RevC – 07/2013

97

88

• The T-piece on the knife holder base (56).

11

10

Knife holder E:
• Clamping lever (11) for the lateral displace-

ment.

• T-piece (88) and guide (97) of the knife holder

for the lateral displacement.

• Clamping lever (10) for the blade.

Knife holders N and NZ:
• Sliding surfaces of the finger guard (8).
• Knurled nuts (73) for measuring height adjust-

ment.

95

96

55

8.2.3 Lubricating the instrument

Once a month, lubricate the following parts with

the included drive parts oil No. 405 (1-2 drops are

enough).

Instrument and specimen holder:
• The clamping key (95) of the quick clamping

system.

• The T-piece (55) on the microtome base plate.
• Guide rails (96) for the knife holder base on the

microtome base plate.

56

8.

Cleaning and Maintenance

73

8

Knife holder N

Knife holder NZ

Fig. 81