Operation – Leica Biosystems RM2265 User Manual

Page 33

Advertising
background image

31

Leica RM2265

Motorized sectioning

In motorized operation, you can choose between four operating modes:
CONT = continuous stroke

SINGLE = single stroke

STEP

= step stroke

PROG = program mode

CONT (continuous stroke) operating mode
• Select CONT operating mode.
After beginning the sectioning process, sectioning continues until the pro-

cess is stopped by pressing RUN/STOP or ENABLE.

The specimen then stops in the next programmed end position of the verti-

cal stroke, depending on which stop position is active.

SINGLE (single stroke) operating mode
• Select SINGLE operating mode.
After starting sectioning, a single sectioning stroke is completed.

The specimen then stops automatically in the programmed end position of

the vertical stroke.

5. Operation

and

In motorized sectioning mode, the sectioning process can be started and stopped with

the foot switch (optional accessory) instead of the RUN/STOP and ENABLE buttons.

For additional information, refer to

Chapter 5, Operation, "Foot switch".

STEP (step stroke) operating mode
• Select STEP (step stroke) operating mode.
After starting the sectioning process, the specimen is moved as long as the

keys are held depressed (or as long as the foot pedal is pressed).

If the buttons or the foot switch are released, the specimen stops automati-

cally.

Advertising