Instrument Setup

The feather key (4) is loosely placed in the hand-

wheel shaft (1) and fixed in place with a cable tie

during transport.
• Remove the cable tie (3).

Caution!

Make sure not to lose the feather key!
• Place the handwheel (2) on the handwheel

shaft (1) as shown.

• Tighten the screw (2a) located in the center

hole of the handwheel with an Allen key size 4.

• Remove the cover foil from the self-adhesive

cover disk and fix the cover disk on the hand-

wheel.

4.4 Assembling the handwheel

4.5 Electrical connections

Checking the voltage
The Leica RM2265 can be connected to various

electrical power grids (depending on voltage and

frequency) and, for this reason, is always deliv-

ered with a set of various power cables.
New instruments are factory-set to 230 volts.

This is documented by a yellow label (230 VOLT)

on the rear of the instrument, which covers the

power switch and power socket.

1

3

2

2a

4

Before connecting the instrument to the

power supply, be absolutely certain to

check that the voltage selector is set to

the voltage in use in your area!

Severe damage can be caused to the

instrument if the voltage selector is set

to an incorrect voltage!

Never change the setting of the voltage

selector while the instrument is con-

nected to the power supply.

The instrument MUST be connected to a grounded power socket. Use only the power

cable provided that matches the power supply (outlets) of the country of use. Do not use

an extension cable without a ground wire!

The handwheel has to be assembled before attempting to use the instrument.

The necessary parts and tools can be found in the toolkit.