5.1 Operating elements and their functions

Three-digit display

for displaying section

thickness/trimming

section thickness.

Four-digit display

of the section counter

MENU MODE + CLEAR

pressed simultaneously

switches to the setting of

the retraction value.

Green LED

Section thickness sum,

Display shows the sum

of all sections in µm.
Green LED

Section counter dis-

plays the number of all

sections.

CLEAR button

Resets the display (sec-

tion counter or section

thickness sum) to 0

MENU MODE button

Switches between

section thickness sum

and section counter.

E-STOP LED

Lights up when emergency

stop function is activated.

M-STOP LED

Lights up when hand-

wheel locking mecha-

nism is activated.

RETRACT LED

Lights up during

specimen retraction.

Green LED

Lights up when trim-

ming mode is activated.
Green LED

Lights up when section-

ing mode is activated.

Fig. 14

Green LED

The vertical stop posi-

tion of the object head

(Up or Lo) is shown in

the display

The operating functions of the microtome are divided between a control panel and a

display unit on the microtome.

A control panel on the instrument displays the current operating mode as well as vari-

ous settings.

All operating functions are centrally located in the separate control panel. All keys and

displays are logically arranged in functional groups and easily identifiable.

5.1.1 Instrument control panel