8 schneiden, 9 probenwechsel bzw. unterbrechung der arbeit, 8 sectioning – Leica Biosystems RM2265 User Manual

Page 45: Changing the specimen or interrupting sectioning, Operation

Advertising
background image

43

Leica RM2265

5.9 Changing the specimen or interrupting sectioning

• Center the handle of the handwheel.

• Use the TRIM/SECT key to select the sectioning mode.

• Adjust the appropriate sectioning thickness or verify the selected value.

• Depending on the operation planned, use the CUT MODE key to select

one of the motorized sectioning modes, CONT, SINGLE or STEP.

5. Operation

5.8 Sectioning

+

Always use a different area of the cutting edge for trimming and sectioning. To do so, laterally

displace the blade or knife in the knife holder.

When using the knife holder E with lateral displacement, it is sufficient to move the knife

holder sideways.

Lock the handwheel and cover the knife edge with the knife guard prior to any manipulation of

knife or specimen head, as well as prior to changing the specimen block and during all work

breaks!

• Raise the specimen to the upper end position and activate the mechani-

cal handwheel lock.

• Cover the sectioning edge with the knife guard.
• Remove the specimen from the specimen clamp and mount a new sam-

ple to continue.

• Before cutting into a new specimen, move the specimen head back to

the rear end position.

Always set the sectioning speed according to the hardness of the

specimen! For hard specimens, always select a slow speed.

• Check the sectioning speed setting and set an appropriate speed.

• Start motorized sectioning.

• Pick up the sections and mount them on microscope slides.

Always turn the handwheel evenly in clockwise direction; other-

wise, the brake will not work properly.