4 universal-kassettenklammer, 4 universal cassette clamp, Optional accessories – Leica Biosystems RM2265 User Manual

Page 53

Advertising
background image

51

Leica RM2265

65

• Push the lever (60) forwards.
• Mount the cassette (65) horizontally or verti-

cally as required.

• To clamp the cassette, release the lever (60).

6.2.4 Universal cassette clamp

60

Order No. 14 0502 37999

Leica / Surgipath cassettes with minimum dimensions of 39.8 x 28 mm and maximum dimensions

of 40.9 x 28.8 mm may be clamped in the universal cassette clamp (UCC) horizontally as well as

vertically.
When using other – particularly thin-walled – cassettes, the cassette may become deformed or

other problems may arise with the clamping system. If the user tries clamping the cassette and

realizes that it is not securely clamped into place, an alternative tensioning clamp must be used.
When using cassettes whose lid is molded on, make sure that the broken edge left by removing

the lid does not prevent the specimen from being securely clamped – if necessary, the specimen

must be clamped horizontally.
Before clamping the cassette into the universal cassette clamp, remove excess wax on the out-

side of the cassette to ensure that the cassette clamps in securely.
Wax deposits on the outside of the cassette can make the universal cassette clamp dirty. The dirt

prevents the cassette from clamping in securely and can lead to sections being too thick or thin,

chattering within the section and, in the worst-case scenario, damage to the specimen. Prior to

sectioning, the user has to verify that the specimen is clamped securely and, if necessary, remove

wax deposits from the universal cassette clamp according to the specifications in

Chapter 8.1

"Cleaning and maintenance – Universal cassette clamp".

Prior to sectioning, laboratory personnel

MUST check that the cassette is seat-

ed securely in the universal cassette

clamp.