Troubleshooting – Leica Biosystems RM2265 User Manual

Page 77

Advertising
background image

75

Leica RM2265

Problem

Possible cause

Corrective action

4. Noises during sectioning
The knife "sings" when section-

ing hard specimens. The sec-

tions show scratches or chatter

marks.

• The sectioning speed is too fast.

• The clearance angle is too

wide.

• Insufficient clamping of speci-

men clamp and/or knife holder.

• Select a slower sectioning

speed.

• Methodically reduce the clear-

ance angle setting until you have

found the optimum angle.

• Check all screw and clamp con-

nections on the specimen holder

system and the knife holder. If

necessary, tighten the levers and

screws.

5. High blade consumption

• Too great of a sectioning force

was applied.

• Adjust the sectioning speed and/

or section thickness when sec-

tioning. Select a thinner section.

7. Troubleshooting