5 schnittabfallwanne, 6 unterflurbeleuchtung, Section waste tray – Leica Biosystems RM2265 User Manual

Page 65: 6 backlighting, Optional accessories

Advertising
background image

63

Leica RM2265

6. Optional Accessories

• Remove the two screws (1) using a slotted

screwdriver and then remove the cover plate

(2).

• Insert the backlighting (3) in the recess at the

rear of the knife holder base.

• Connect the plug (4) of the backlighting to the

socket (5) on the microtome.

The backlighting illuminates once the microtome

is turned on with the power switch.

2

1

3

4

1

5

6.6 Backlighting

• Push the section waste tray (18) from the front

to the microtome base plate (53) until it is held

in place by the two magnets (39) (on the front

of the microtome base plate).

To remove the section waste tray, lift it slightly

and pull it off towards the back.

6.5 Section waste tray

53

18

39

NEVER use the backlighting illumi-

nation with the super mega-cassette

clamp!

The backlighting is inserted into the

front of the one-piece knife holder

base.

Fig. 55

Fig. 56

Advertising