6 präparathalter, 7 präparathalteraufnahme, 6 specimen holders – Leica Biosystems RM2265 User Manual

Page 55: 7 specimen holder fixture, Optional accessories

Advertising
background image

53

Leica RM2265

6. Optional Accessories

1

3

2

3

5

1

4

2

6.2.6 Specimen holders

The specimen holders are especially intended

for small specimens. They are available in two

models: as an EM flat specimen holder (

Fig. 41

, 4)

and an EM universal specimen holder (

Fig. 40

, 1).

You can insert either specimen holder into the

segment arc or the specimen holder fixture.

Clamping the specimens
• Insert the special key (2) into the clamping

screw (3) of the specimen holder and open

the clamping chucks of the specimen holder

by turning the special wrench.

• Insert the specimen.

• Clamp the specimen by turning the special

wrench.

6.2.7 Specimen holder fixture

The specimen holder fixture (1) is inserted into

the dovetail fixture in the same way as a speci-

men clamp.
• Open the knurled screw (2) by turning it coun-

terclockwise.

• Insert the pin (3) of the specimen holder (4)

into the hole (5).

• Turn the knurled screw (2) clockwise to clamp

the specimen holder.

If the backlighting illumination is

used, ensure that the clamping screw

is pointing up.