5. Operation

There are two versions of the specimen holder

fixture, one with and one without specimen ori-

entation, which are interchangeable.

The specimen orientation allows for simple posi-

tion correction of the specimen surface when the

specimen is clamped into place.

You can use the specimen holder fixture (60) to

hold all available accessory specimen clamps

(for additional information, refer to

Chapter 6

"Optional Accessories"

).

To do so, proceed as follows:

• Move the specimen holder fixture (60) to the

upper end position by turning the handwheel

(12) and engage the handwheel lock.

• To release the clamping system, turn the

screw (61) counterclockwise using an Allen

key size 4 (71).

• Push the guide (63) of the standard clamp (62)

from the left into the specimen holder fixture

(60) as far as it will go.

• To clamp the specimen clamp, tighten the

screw (61) in a clockwise direction using an

Allen key size 4 (71).