Instrument components and specifications – Leica Biosystems RM2265 User Manual

Page 16

Advertising
background image

14

Instructions for Use, V 1.8 RevC – 07/2013

3.

Instrument components and specifications

Microtome

Sectioning thickness setting:
Section thickness setting range:

0.25 - 100 µm

Setting values:

0.25 µm

0.50 - 5.0 µm

in 0.5 µm increments

5.0 - 20.0 µm

in 1.0 µm increments

20.0 - 60.0 µm

in 5.0 µm increments

60.0 - 100.0 µm

in 10.0 µm increments

Trimming section thickness setting range: 1 - 600 µm
Setting values:

1.0 - 10.0 µm

in 1.0 µm increments,

10.0 - 20.0 µm

in 2.0 µm increments,

20.0 - 50.0 µm

in 5.0 µm increments,

50.0 - 100.0 µm

in 10.0 µm increments,

100.0 - 600.0 µm

in 50.0 µm increments.

Specimen feed:

28 mm ±1 mm, feed motion via step motor

Vertical stroke:

70 mm

Maximum sectioning area w/o retraction: 65 mm without specimen orientation
Maximum sectioning area with retraction: 60 mm
Specimen retraction:
in manual sectioning mode:

5 - 100 µm in 5 µm increments; can be turned off

in motorized sectioning mode:

Varies with the sectioning speed; can be turned off

Electric coarse feed:

300 µm/s and 900 µm/s

Sectioning speed:

0.5 - 420 mm/s

±10 %

Return speed

approx. 120 - 420 mm/s ±10 %

Repositioning of knife holder base
North-south:

24 mm

Maximum specimen size (L x H x W):

50 x 60 x 40 mm

Specimen orientation
Horizontal:

Vertical:

Advertising