Troubleshooting – Leica Biosystems RM2265 User Manual

Page 76

Advertising
background image

74

Instructions for Use, V 1.8 RevC – 07/2013

7. Troubleshooting

Problem

Possible cause

Corrective action

• The blade is dull.

• The pressure plate is damaged

or incorrectly adjusted.

• Clearance angle of the knife/

blade too small.

• When using cassettes, par-

ticularly thin-walled cassettes,

from a company other than

Leica or Surgipath, the cassette

may become deformed or other

clamping problems may arise.

If while trying to clamp the cas-

sette you realize that it is not

securely clamped into place,

an alternative tensioning clamp

must be used.

• Laterally displace the knife

holder or insert a new blade.

• Insert a new pressure plate or

use a new knife holder.

• Methodically experiment with

larger clearance angle settings

until you have found the optimum

angle.

2. Section compression
The sections are very compressed,

show folds or are squeezed

together.

• The blade is dull.

• The specimen is too warm.

• The sectioning speed is too

fast.

• Use another area of the blade or

a new blade.

• Cool the specimen before sec-

tioning.

• Reduce the sectioning speed.

3. "Stripes" in sections
For knife holder E

• There is an accumulation of par-

affin on the rear pressure plate

of the knife holder.

• Remove paraffin from this area

on a regular basis.