7 anschneiden der probe, 1 anschneiden im manuellen schneidebetrieb, 2 anschneiden im motorischen schneidebetrieb – Leica Biosystems RM2265 User Manual

Page 44: Trimming the specimen, 1 trimming in manual operation, 2 trimming in motorized operation, Operation

Advertising
background image

42

Instructions for Use, V 1.8 RevC – 07/2013

• Using the rotary knob, set the appropriate sectioning speed.
• Using the CUT MODE button, select the CONT operating mode (continu-

ous stoke).

• Deactivate the handwheel lock and release the brake.
• Start motorized sectioning and trim the sample.
• Terminate trimming when the desired sectioning surface and depth have

been reached.

5. Operation

5.7 Trimming the specimen

5.7.1 Trimming in manual operation

• Use the TRIM/SECT key to select the trim mode.
• Set the desired trim section thickness.
• Deactivate the handwheel lock and release the brake.
• In TRIM mode, use the coarse feed buttons to move the specimen

against the knife/blade.

• Trim the specimen by turning the handwheel

or

• Using the CUT MODE key, select the ROCK operating mode and cut the

specimen with forward and backward motions of the handwheel.

• Terminate trimming when the desired sectioning surface and depth have

been reached.

5.7.2 Trimming in motorized operation

• Use the TRIM/SECT key to select the trim mode.
• Set the desired trim section thickness.
• If necessary, set the sectioning window.

During fast manual trimming, do not put your fingers between the

specimen and knife. The handwheel continues turning after it is

released and can cause injuries.

Always set the sectioning speed according to the hardness of the

specimen! For hard specimens, always select a slow speed.

The handle of the handwheel must always be centered while in mo-

torized sectioning mode.

+

Advertising