47

Leica RM2265

Assembly of the super mega-cassette clamp

The super mega cassette clamp should preferably be used with the rigid

fixture for specimen clamps.

To do so, proceed as follows:

6.1.5 Super mega-cassette clamp

Fig. 31

1

2

3

4

5

6. Optional Accessories

If the directional fixture for specimen clamps is used with the rigid knife holder base,

the orientation must be in position "0" and the cover for backlighting illumination must

be detached. (Danger of collision if not observed!)

NEVER use the super mega-cassette clamp with backlighting illumination!

• Screw the rigid fixture for specimen clamps (4) onto the specimen

head (3):

Remove the screw (1), place the specimen holder fixture (4) onto the

specimen head (3) from the front and tighten the screws (2) with an Allen

key No. 3.

Next, insert the screw (1) from the side and briefly tighten it with an Allen

key No. 4.

• Insert the super mega-cassette clamp from the side on the left into the

dovetail guide of the rigid fixture for specimen clamps and tighten screw (1).

Remove the rubber ring (5) only after having attached the cassette

clamp on the specimen head!