Operation – Leica Biosystems RM2265 User Manual

Page 43

Advertising
background image

41

Leica RM2265

• Carefully push in the blade from above or from

the side. Make sure that the blade is clamped

in the center and, most importantly, parallel to

the upper edge of the pressure plate.

• To clamp the blade, rotate clamping lever (10)

clockwise back upwards.

Fig. 28

10

10

9a

9

• To remove the blade, fold the clamping lever

(10) downwards counterclockwise.

The ejector (9a) ensures safe blade changing.

Once the blade has been removed from the blade

holder, it is disposed of into the dispenser con-

tainer (underside, see image).

• Another option for removing the blade is to use

the brush with magnet.

To do so, fold the clamping lever (10) down-

wards counterclockwise. Likewise, fold the

knife guard (9) downward. Guide the brush

with magnet to the blade and lift it upwards

and out.

9

5. Operation

Use the blade ejector to eject the blade!

Advertising