2.3 Integrated safety devices

2. Safety

Handwheel lock (manual operation only)
• To lock the handwheel, push the lever (5) outwards and

continue to turn the handwheel slowly until it locks exactly

in the 12 o'clock position. The LED (4) in the LOCK display

lights up.

Handwheel brake
Using the lever (3) on the right side of the microtome base

plate, the handwheel can be braked in any position.

• To lock the handwheel, pull the lever forward forcefully.

• To unlock the handwheel brake, push the locking lever (3)

back to its original position.

Emergency-stop function
The emergency-stop function is activated with the red

EMERGENCY STOP switch (1) on the upper left of the front

of the microtome. The sectioning motor stops immediately

when the EMERGENCY STOP switch is pressed. The red LED

in the E-STOP field (2) on the control panel of the instrument

lights up, indicating that the emergency stop function has

been activated.

To deactivate this function, turn the EMERGENCY STOP

switch in the direction of the arrow.

3

1

4

2

5

2

12

Fig. 2

Fig. 3

Important!

The LED (4) in the M-STOP field indicates only that the instrument cannot be started. It does not

provide any indication that the handwheel brake is active.

The lever (3) must be pulled completely forward with force so that the handwheel brake is

applied. The handwheel is securely locked only when the lever (5) is in the 12 o'clock position.

Caution!

Never operate the lever (5) during motorized

sectioning.